Yogi Prabowo
"Reseksi tulang distal femur memerlukan tindakan rekonstruksi defek osteoartikular menggunakan megaprostesis distal femur. Rekonstruksi megaprostesis memiliki masa pakai terbatas yang membutuhkan tindakan pergantian atau revisi. Tindakan revisi megaprostesis menimbulkan kehilangan masa tulang dan membutuhkan ukuran megaprostesis yang lebih panjang. Belum ada megaprostesis dengan ruang tandur tulang yang dapat memfasilitasi regenerasi tulang.
Penelitian ini mengembangkan desain prototipe megaprostesis dengan metode reverse engineering menggunakan CT scan femur orang dewasa dan implan megaprostesis benchmarking, melakukan pengujian simulasi Finite Element Analysis (FEA) dan pengujian biomekanik menggunakan mesin uji Universal Tensile Machine (UTM) Tensilon RTF-2350 berdasarkan skema ISO 10328 dan ASTM F1800, analisis antropometri ekstremitas bawah dan lutut pada anak remaja dan dewasa dan implementasi pada kadaver untuk mendeskripsikan langkah pembedahan dan analisis inter-observer diskusi pakar tentang kelayakan pemasangan prototipe.
Prototipe α plus megaprostesis femur distal dengan ruang tandur tulang dibuat dari bahan stainless steel dan polyethylene dengan menggunakan mesin Computer Numerical Control/CNC 3 dan 5 aksis. Prototipe ini adaptif secara modular dengan paku intrameduler femur (femoral intramedullary nail) dan pelat pengunci (broad plate locking). Kombinasi ini memberi ruang tandur tulang untuk menerapkan tehnik Masquelet dalam regenerasi tulang. Pada simulasi Finite Elemen Analysis (FEA) dan pengujian biomekanik didapatkan titik lemah pada sambungan ujung distal paku intrameduler femur dengan sekrup pengunci (locking screw) , femoral block step (FBS) 1 dan 2 yang terbuat dari polyethylene dan sambungan pelat pengunci dengan sekrup pada femur proksimal. Terdapat perbedaan antropometri ekstremitas bawah dan sendi lutut antara anak remaja dan dewasa, laki laki dan perempuan yang memberikan pertimbangan pembuatan ukuran implan yang berbeda. Tehnik pembedahan dan pemasangan implan dapat diimplementasi pada kadaver dan lebih dari 60 % pakar ortopedi onkologi memberikan opini sangat baik terhadap kelayakan pembedahan dan pemasangan implan prototipe.
Prototipe α plus megaprostesis femur distal dengan ruang tandur tulang berpotensi memberikan solusi dalam hal ruang untuk regenerasi tulang dengan penyempurnaan desain dan pemilihan bahan manufaktur implan yang lebih baik untuk diterapkan pada manusia.

Resection for femur distal bone requires the reconstruction of osteoarticular defects using distal femur megaprostheses. Megaprosthesis reconstruction has a limited lifespan that requires further replacement or revision. The revision of the megaprosthesis causes bone loss and requires a longer megaprosthesis size. There is no megaprosthesis with a bone chamber that can facilitate bone regeneration.
This study develop a prototype design of a megaprosthesis by reverse engineering method using CT scan of adult femur and benchmarking megaprosthesis implant, conducted Finite Element Analysis (FEA) simulation testing and biomechanical testing using the Tensilon RTF-2350 Universal Tensile Machine (UTM) testing machine based on ISO 10328 and ASTM F1800 schemes, anthropometric analysis of lower extremities and knees in adolescents and adults and implementation on cadavers to describe surgical steps and inter-observer analysis expert discussion on the feasibility of prototyping.
The prototype α plus distal femur megaprosthesis with a bone chamber is made from stainless steel and polyethylene materials using 3 and 5 axis Computer Numerical Control/CNC machines. The prototype is modularly adaptive with femoral intramedullary nails and broad plate locking. This combination gives the bone chamber to apply the Masquelet technique in bone regeneration. Simulation of Element Analysis (FEA) and biomechanical testing show weakness points were obtained at the distal end joints of intramedullary femoral nail with locking screw, femoral block steps (FBS) 1 and 2 which made of polyethylene, and the connection of the locking plate with screws on the proximal femur was obtained. There are differences in the anthropometry of the lower extremities and knee joints between adolescents and adults, men and women, which gives consideration to the manufacture of different implant sizes. Surgical techniques and implant placement can be implemented on the cadaver and more than 60% of orthopedic oncology experts give a very good opinion on the feasibility of surgery and installation of prototype implants.
The prototype of α plus distal femur megaprosthesis with bone plantation chamber has the potential to provide a solution in terms of space for bone regeneration with improved design and better selection of implant manufacturing materials for application in humans.
"
Jakarta: Fakultas Kedokteran Universitas Indonesia, 2025
D-pdf
UI - Disertasi Membership  Universitas Indonesia Library
cover
Muhammad Sjahrul Annas
"ABSTRAK
Sendi lutut adalah persendian yang amat dominan perannya dalam tubuh kita. Sendi ini dapat mengakomodasi tegangan yang amat tinggi sekaligus tetap dapat mempertahanakan tingkat mobilitas dan stabilitasnya Bentuknya yang spesifik dan ukuran sambungan yang terbesar dalam tubuh memperlihatkan kemampuannya untuk melakukan fungsi diatas. Beban yang diterima oleh lutut minimal 3 kali berat badan penggunanya dalam keadaan jalan dan 5 kali pada saat menaiki tangga. Sungguh suatu ciptaan Allah yang sangat jenius. Berdasarkan fungsi dan beban lutut, maka lutut merupakan bagian yang sangat penting untuk mobilitas manusia. Studi ini bertujuan untuk mencari konsep perancangan lutut buatan yang dapat mempertemukan aspek teknis dan operasi produk serta aspek konsumen.
Perancangan ini berdasarkan pada kehendak pemakai berdasarkan survei yang dilakukan pada pengguna dan pemakai lutut buatan. Hasil perancangan ini, lutut buatan yang terbaik adalah lutut buatan yang bentuknya sesuai dengan bentuk lutut sesungguhnya.
Hasil rancangan ini menjadi masukan bagi pengembangan selanjutnya untuk menghasillkan lutut buatan ideal yang diinginkan.

ABSTRACT
Knee joint is a very dominant role in our body. This joint not only can accomodate a very high pressure but also keep the mobility and stability as well. The specific shape and the biggest joint size in the body, shows the performance to do the function. The load it got is at least 3 times of the weight of the walking body and 5 times of a body climbing a stairs. It really is a very genius God's creation. Based on the knee function and load, knee is the most important part of human mobility.
The objective of this study is to find an artificial knee design that meets technical aspect, operational product and customer need. This design based on customer needs surveys from users.
The result is: the best artificial knee is the one that has the same shape as the real knee. This become an input for the next developments to have the ideal artificial knee.
"
Depok: Fakultas Teknik Universitas Indonesia, 1998
T-Pdf