"ABSTRAK
Pasien spesifik QA pada teknik IMRT bertujuan untuk menjamin bahwa parameter-parameter penyinaran sesuai dengan perencanaan terapi. Dikarenakan terjadinya setup error pada pengukuran pasien spesifik QA pada teknik IMRT cukup tinggi, sehingga in house program QA akan berguna sebagai QA alternatif. Penelitian dilakukan menggunakan pesawat Varian Unique milik Rumah Sakit Cipto Mangunkusumo RSCM . Dynalogfile diperoleh dari tujuh kasus penyinaran. QA MLC dianalisa melalui parameter root mean square RMS error. Sementara itu, fluence map direkonstruksi berdasarkan informasi dari posisi leaf, posisi jaw, dan fraksional MU yang diperoleh berdasarkan data Dynalogfile setiap 50 ms. Tingkat konsistensi fluence map prediksi antar fraksi dievaluasi menggunakan gamma index. Kalkulasi gamma index antara fluence map sebenarnya dan fluence map prediksi dilakukuan untuk mengevaluasi adanya potensi error dari segmen-segmen yang dibentuk oleh posisi MLC. Kalkulasi dosis dilakukan berdasarkan data fluence map menggunakan metode modified Clarkson integration MCI . Hasil QA MLC pada seluruh kasus adalah kurang dari 0.5 mm untuk mean RMS error dan kurang dari 1 mm untuk max RMS error. Hasil ini menunjukkan bahwa posisi MLC saat penyinaran sesuai dengan perencanaan terapi. Evaluasi gamma dari fluence map sebenarnya terhadap fluence map prediksi pada pada seluruh sampel memiliki nilai rata-rata gamma dengan kriteria 2 , 2mm adalah 97.6 1.3 , sedangkan pada kriteria 3 , 3mm adalah 99.0 0.6 . Hasil ini menunjukkan bahwa tidak terdapat signifikan error pada segmen-segmen lapangan IMRT. Kalkulasi dosis menggunakan modified Clarkson Integration MCI berdasarkan fluens MU dibandingkan dengan kalkulasi dosis pada perencanaan terapi secara umum menghasilkan nilai evaluasi gamma dengan kriteria 4 ,4mm adalah lebih dari 95 pada lapangan kecil non split IMRT dan memiliki nilai lebih dari 90 untuk lapangan besar split IMRT .

ABSTRACT
The purpose of IMRT patient specific QA is to ensure that delivery radiation parameters conform with treatment planning system. Since there are high probability of setup error in measurement IMRT patient specific QA, the in house QA program is useful as an alternative QA. This study aimed to investigate MLC QA and patient specific QA based on Dynalogfile. Varian Linear Accelerator at Cipto Mangunkusumo hospital have been used for this study. Dynalogfile were obtained from seven cases. The MLC QA were analyzed using the error RMS leaf parameter. Meanwhile, the fluence map were reconstructed based on the information of leaf position, jaw position, and fractional MU from Dynalogfile data every 50 ms. The consistency check of expected fluence map from fraction to fraction were calculated using gamma index evaluation. The gamma index evaluation between actual and expected fluence map were calculated to analyze the segment potential error caused by MLC positions. The dose were calculated using modified Clarkson integration based on fluence map data. The results of MLC QA from all cases are less than 0.5 mm for mean RMS error and below 1 mm for max RMS error. These results indicate that actual MLC position conform with treatment planning system. The results of 2 ,2mm and 3 , 3mm gamma evaluation from all samples between actual and expected fluence map are 97.6 1.3 and 99.0 0.6 . Those indicated that there is no IMRT segment significant error. Gamma pass rate results of 4 , 4 mm criteria between dose calculation using modified Clarkson integration based on fluence map and TPS calculation are above 95 for small IMRT field non split IMRT and above 90 for large IMRT field split IMRT . "
2018
T51638
UI - Tesis Membership  Universitas Indonesia Library
cover
Wahyu Edy Wibowo
"[Telah dilakukan penelitian dalam mengevaluasi pengaruh material homogen dan inhomogen pada teknik IMRT (Intensity Modulated Radiotherapy) menggunakan modalitas Fan Beam CT (FBCT) dan Cone Beam CT (CBCT) terhadap Gamma Index (GI). Perencanaan ulang dilakukan untuk 5 pasien kanker paru kanan pada modalitas FBCT dan CBCT menggunakan TPS Phillips Pinnacle energi foton 6 MV. Menggunakan Teknik IMRT 30 segmen, 50 segmen, dan 70 segmen untuk perencanan terapi. Calculation Grid Resolution (CGR) 0.2 cm dan 0.4 cm digunakan sebagai resolusi dalam perhitungan dosis. Serta kalibrasi bilangan CT (KBC) FBCT, CBCT, dan linear dipergunakan sebagai koreksi CT density number. Pengukuran plannar dose untuk evaluasi Gamma Index (DD 2% / DTA 2 mm, passing rate 90%) dilakukan pada SAD 100 cm dengan material homogen dan inhomogen pada kedalaman 1.5 cm, 5 cm, dan 10 cm. Didapatkan deviasi nilai rata-­‐rata GI antara CGR 0.2 cm dengan 0.4 cm material homogen modalitas FBCT dan CBCT berturut-­‐tutut ≤1.46% dan ≤ 1.13%. Serta deviasi pada material inhomogen sebesar ≤ 2.54% (FBCT) dan ≤ 1.74% (CBCT). Deviasi Nilai rata-­rata GI antara FBCT dengan CBCT dengan CGR 0.2 cm berturut-­turut ≤ 1.95% (homogen) dan ≤ 2.36% (inhomogen). Dan deviasi Nilai rata-­‐rata GI antara material homogen dan material ekuivalen homogen sebesar ≤ 0.80%. Dari hasil penelitian yang didapat, bahwa evaluasi Gamma Index dapat dipengaruhi oleh jumlah segmen, modalitas CT, kalibrasi bilangan CT, Calculation Grid Resolution, dan kedalaman material.

These Studies have been carried out to evaluate the effect of a homogeneous and inhomogeneous material on IMRT Technique (Intensity Modulated Radiotherapy) using Fan Beam CT (FBCT) and Cone Beam CT (CBCT) for the Gamma Index (GI). The Phillips Pinnacle treatment plan was used to replan 5 patients on right side of lung cancer. Photon 6 MV was applied to this technique with 30 segments, 50 segments, and 70 segments, repectively. Using Calculation Grid Resolution (CGR) 0.2 cm and 0.4 cm for resolution in the calculation of the dose. As well as the calibration of CT numbers (CCN) FBCT, CBCT, and the linear density are used as a correction CT number. The planar measurement for the evaluation of Gamma Index (DD 2% / DTA 2 mm, the passing rate of 90%) carried out at 100 cm SAD with the homogeneous and inhomogeneous material at a depth of 1.5 cm, 5 cm, and 10 cm, respectively. We obtained deviation average value of GI between CGR 0.2 cm and 0.4 cm using the homogeneous material on FBCT and CBCT modality ≤1.46% and ≤ 1:13%, respectively. As well as the material inhomogeneous deviation of ≤ 2.54% (FBCT) and ≤ 1.74% (CBCT). The deviation of the average GI value between FBCT with CBCT with CGR 0.2 cm respectively ≤ 1.95% (homogeneous) and ≤ 2.36% (inhomogeneous). Finally, the deviation of the average GI value between homogeneous and equivalent of homogeneous material is ≤ 0.80%. From the results obtained, that the Gamma Index can be influenced by the number of segments, modalities of CT, calibration of CT numbers, calculation grid resolution and depth of material.
, These
Studies
have
been
carried
out
to
evaluate
the
effect
of
a
homogeneous
and
inhomogeneous
material
on
IMRT
Technique
(Intensity
Modulated
Radiotherapy)
using
Fan
Beam
CT
(FBCT)
and
Cone
Beam
CT
(CBCT)
for
the
Gamma
Index
(GI).
The
Phillips
Pinnacle
treatment
plan
was
used
to
replan
5
patients
on
right
side
of
lung
cancer.
Photon
6
MV
was
applied
to
this
technique
with
30
segments,
50
segments,
and
70
segments,
repectively.
Using
Calculation
Grid
Resolution
(CGR)
0.2
cm
and
0.4
cm
for
resolution
in
the
calculation
of
the
dose.
As
well
as
the
calibration
of
CT
numbers
(CCN)
FBCT,
CBCT,
and
the
linear
density
are
used
as
a
correction
CT
number.
The
planar
measurement
for
the
evaluation
of
Gamma
Index
(DD
2%
/
DTA
2
mm,
the
passing
rate
of
90%)
carried
out
at
100
cm
SAD
with
the
homogeneous
and
inhomogeneous
material
at
a
depth
of
1.5
cm,
5
cm,
and
10
cm,
respectively.
We
obtained
deviation
average
value
of
GI
between
CGR
0.2
cm
and
0.4
cm
using
the
homogeneous
material
on
FBCT
and
CBCT
modality
≤1.46%
and
1:13%,
respectively.
As
well
as
the
material
inhomogeneous
deviation
of
2.54%
(FBCT)
and
1.74%
(CBCT).
The
deviation
of
the
average
GI
value
between
FBCT
with
CBCT
with
CGR
0.2
cm
respectively
1.95%
(homogeneous)
and
2.36%
(inhomogeneous).
Finally,
the
deviation
of
the
average
GI
value
between
homogeneous
and
equivalent
of
homogeneous
material
is
0.80%.
From
the
results
obtained,
that
the
Gamma
Index
can
be
influenced
by
the
number
of
segments,
modalities
of
CT,
calibration
of
CT
numbers,
calculation
grid
resolution
and
depth
of
material.
Key
words
:
Calculation
Grid
Resolution,
Calibration
of
CT
Numbers,
Cone]
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2016
T45207
UI - Tesis Membership  Universitas Indonesia Library
cover
Wisnu Wardana
"Penelitian ini dimotivasi oleh mesin multileaf collimator (MLC) yang berfungsi untuk mendistribusikan dosis-dosis radiasi yang dihimpun dalam matriks dosis pada pengembangan metode pengobatan kanker dengan radioterapi, yaitu Intensity Modulated Radiation Therapy (IMRT). MLC tidak bisa mengirimkan semua bentuk matriks dosis, sehingga perlu dilakukan suatu dekomposisi matriks agar MLC bisa mengirimkan dosis radiasi. Oleh karena itu, penelitian ini bertujuan untuk mencari cara mendekompsisikan matriks dosis yang dapat dikirim oleh MLC serta mengoptimalkan pendistribusian dosis radiasi. Algoritma Greedy dianggap sebagai algoritma yang paling optimal untuk mendekomposisikan matriks dosis menjadi matriks yang dapat didistribusikan oleh MLC. Akan tetapi ada kasus dimana Algoritma Greedy memberikan hasil yang kurang optimal, sehingga penulis mencoba untuk memodifikasi Algoritma Greedy yang menghasilkan dekomposisi yang lebih optimal.

This research was motivated by the multileaf collimator (MLC) machine, which serves to distribute the doses of radiation that is collected in the matrix dosage in the development of treating cancer methods with radiotherapy that is Intensity Modulated Radiation Therapy (IMRT). MLC cannot send all form of matrix dosage, so we need some matrix decomposition that MLC can transmit the radiation dose. Therefore, this study aimed to find the ways how to decompose matrix dosage that can be delivered by the MLC and optimize the distribution of radiation dose. Greedy algorithms are considered as the most optimal algorithm for decomposing the matrix into matrix dosage that can be distributed by the MLC. There are some cases where the Greedy algorithms provide the less optimal results. Thus in this research the algorithm is modified to obtain the more optimal result.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2017
S-69123
UI - Skripsi Membership  Universitas Indonesia Library
cover
Pamuji Widodo
"Verifikasi dosis harus dilakukan pada setiap pra-terapi yang menggunakan teknik IMRT. Salah satu alat ukur dosimetri IMRT adalah matriks jenis ion chamber matriXX-FFF,IBA yang memberikan hasil pengukuran dosis secara real-time, serta tidak diperlukan kalibrasi berkala. Tetapi, dosimetri tersebut memiliki resolusi spasial terbatas, tergantung pada ukuran dan bentuk detektor serta ada efek transport elektron sekunder dari dinding detektor ke dalam volume detektor, sehingga seringkali mengakibatkan hasil verifikasi cukup rendah. Hal tersebut salah satunya dapat diatasi dengan metode back projection yang diusulkan oleh Poppe et al. Pada penelitian ini metode tersebut diterapkan dengan menggunakan fungsi kernel dari lapangan tunggal, kemudian diaplikasikan pada variasi lapangan persegi, kondisi No-BuildUp NBU dan BuildUp BU, posisi Extended-SSD, serta kasus IMRT dengan energi 6 MV dan 10 MV. Analisis indeks gamma Dose Difference/DD dan Dose to Agreement/DTA digunakan untuk memverifikasi efektivitas metode ini. Fungsi kernel diperoleh dari lapangan persegi 4 cm dengan dekonvolusi perbandingan MatriXX-FFF dan TPS. Hasil menunjukan terjadi peningkatan nilai passing grade secara keseluruhan 0 - 15.9, = 4.84, ? = 4.81, dimana nilai tertinggi pada lapangan persegi 4 cm dan 5 cm, baik kondisi NBU dan BU. Hal tersebut mengindikasikan adanya ketergantungan kernel terhadap ukuran lapangan. Hasil penerapan pada kasus IMRT juga memberikan kecenderungan peningkatan nilai passing grade. Kedepannya perlu dilakukan penelitian yang lebih lanjut mengenai hubungan fungsi kernel terhadap ukuran lapangan serta penerapan pada kasus IMRT lebih banyak.

Verification process has to be carried out every pre treatment by using Intensity Modulated Radiotherapy IMRT . One of dosimetry tools for IMRT is matrix detector type of ionization chamber that gives the result of dose measurement in real time, and is not required calibration periodically. But, the dosimetry has finite spatial resolution, depends on size and shape of detector, and occur the effect of secondary electron transpot from detector wall to detector volume, so that often gives the result of gamma parameter low enough. One of solutions, it can be solved by back projection method proposed by Poppe et al. In this research, that method could be applied by using kernel function of single field, then be used for variations of square field, no BuildUp NBU and BuildUp BU condition, extended SSD position, and IMRT case with 6 MV and 10 MV. Analysis of gamma parameter Dose Difference DD and Dose to Agreement DTA is used to verify the effectiveness of this method. The kernel function is obtained from square field of 4 cm by deconvolution from comparison of MatriXX FFF and TPS. The results show increasing passing grade value for all of fields 0 15.9, 4.84, 4.81 , where the highest values are on square field of 4 cm and 5 cm at NBU and BU condition. All of those indicate the dependence kernel function to size of field. Application result of IMRT cases also give the same trend of increasing passing grade value. For the next, it needs to do research more about the relationship kernel function to size of field and the application of this method on IMRT case.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2017
T49122
UI - Tesis Membership  Universitas Indonesia Library
cover
Vika Vernanda
"

Salah satu kekurangan teknik Intensity Modulated Radiation Therapy (IMRT) adalah dosis serap di jaringan sehat yang cukup tinggi. Berkas proton memiliki karakteristik yang mampu mengkompensasi kekurangan tersebut. Karakteristik bragg peak yang dimiliki berkas proton memungkinkan dosis tinggi hanya pada target. Kasus Non-Small Cell Lung Cancer (NSCLC) terletak di sekitar banyak organ vital, sehingga deposisi dosis yang melebihi batas akan berdampak signifikan. Proton juga merupakan partikel bermassa yang menunjukkan pola interaksi dengan heterogenitas jaringan yang berbeda dengan foton. Penelitian ini bertujuan untuk mengetahui distribusi dosis perencanaan berkas proton pada kasus NSCLC dengan teknik IMPT serta membandingkan efektivitasnya dengan teknik IMRT. Perencanaan dilakukan menggunakan TPS Eclipse pada fantom air dan citra fantom in-house thorax dynamic. Perencanaan pada fantom air menggunakan 1 lapangan pada 0o dan 3 lapangan pada 45o, 135o, dan 225o. Perencanaan pada fantom in-house thorax dynamic dilakukan menggunakan single field, sum-field, dan multiple field. Nilai Conformity Index (CI) dan Homogeneity Index (HI) antara perencanaan IMPT dan IMRT tidak menunjukkan perbedaan yang signifikan. Gradient Index (GI) perencanaan IMPT berkisar antara 4,15-4,53, sedangkan nilai GI perencanan IMRT sebesar 7,89. Terdapat perbedaan yang cukup signifikan antara keduanya. Histogram distribusi dosis planar menunjukkan bahwa hasil perencanaan IMPT memberikan dosis rendah di luar target lebih sedikit dibandingkan hasil perencanaan IMRT. Selain itu dilakukan juga pengukuran dosis hasil perencanaan IMPT pada lima posisi target, serta empat posisi OAR. Hasilnya dibandingkan dengan data pengukuran IMRT. Nilai dosis titik pada target tidak berbeda secara signifikan, namun nilai dosis empat posisi OAR adalah nol, yang menunjukkan reduksi signifikan dibandingkan teknik IMRT.


One deficiency of the Intensity Modulated Radiation Therapy (IMRT) technique is that the absorbed dose in healthy tissue is quite high. Proton beams has characteristics that can compensate for these deficiencies. The bragg peak characteristic of a proton beam allows high doses only to the target. Non-Small Cell Lung Cancer (NSCLC) cases are located around many vital organs, so the doses that exceed the limit will has a significant impact. Protons are also heavy particles that show patterns of interaction with tissue heterogeneity, which is different with photons. This study aims to determine the dose distribution of proton beam planning in the NSCLC case with the IMPT technique and to compare its effectiveness with the IMRT technique. Planning was done by using TPS Eclipse on the water phantom and in-house thorax dynamic phantom. Planning parameter on the water phantom used 1 field at 0o and 3 fields at 45o, 135o, and 225o. Moreover, we used single field, sum-field, and multiple fields techniques on the in-house thorax dynamic phantom. Conformity Index (CI) and Homogeneity Index (HI) showed a bit differences between IMPT and IMRT planning. The Gradient Index (GI) of IMPT planning ranges between 4.15-4.53, while the GI value of IMRT is 7.89. The planar dose distribution histogram showed that the results of IMPT planning gave fewer out of target doses than IMRT planning results. In addition, evaluation was also made on the target of IMPT planning at five area of interest, as well as four OAR positions. The results are compared with IMRT measurement data. The point dose value at the target did not differ significantly, however the absorbed dose of the four OAR positions are zero which had a large deviation compared to the IMRT technique.

"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2020
S-pdf
UI - Skripsi Membership  Universitas Indonesia Library
cover
Meirisa Ambalinggi
"Intensity modulated radiation therapy (IMRT) memberikan peningkatan kemampuan untuk menyesuaikan distribusi isodose dengan bentuk target sehingga mengurangi dosis ke organ-at risk. Namun kemampuan IMRT ini disertai oleh kompleksitas lapangan penyinaan IMRT sehingga direkomendasikan untuk melakukan patient-spesific pretreatment quality assurance. QA yang umum digunakan adalah QA berbasis pengukuran, tetapi ini merupakan prosedur yang memakan waktu. Sehingga dibuat suatu alat ukur dari proses jaminan kualitas yang disebut metrik kompleksitas. Studi ini bertujuan untuk mengevaluasi metrik kompleksitas berbasis bukaan pada bukaan multileaf collimator (MLC) yang digunakan dalam klinis dan menganalisis korelasinya dengan gamma pass rate (GPR). Nilai metrik kompleksitas dihitung menggunakan Matlab. Ionization chamber 2D array MatriXX Evolution digunakan untuk verifikasi dosis. Distribusi dosis 2D dianalisis dengan software OmniPro-I'mRT dan dibandingkan menggunakan kriteria indeks gamma 3%/3mm dan 3%/2mm dengan low dose threshold 10%. Hasil uji korelasi metrik kompleksitas dan GPR menunjukkan EAM memiliki korelasi yang baik dengan GPR untuk kedua kasus. Nilai koefisien korelasi untuk kasus CNS adalah –0,918 (3%/3mm) dan –0,864 (3%/2mm), sedangkan untuk kasus payudara nilai koefisien korelasinya adalah –0,983 (3%/3mm) dan –0,961 (3%/2mm). Sedangkan untuk MCS, CAM, CPA, dan MU/Gy memiliki korelasi yang lemah dengan nilai GPR.

Intensity-modulated radiation therapy (IMRT) provides an increased ability to adjust the isodose distribution to the target shape, thereby reducing the dose to organ-at-risk. However, IMRT’s capabilities come with the complexity of IMRT’s treatment field, so it is recommended to carry out patient-specific pretreatment quality assurance. The commonly used QA is measurement-based; however, it is time-consuming. So that made a measuring instrument of the quality assurance process called the complexity metric. This study aims to evaluate aperture-based complexity metrics for multileaf collimator (MLC) apertures used in clinical practice and analyze its correlation with gamma pass rate (GPR). Complexity metric are calculated using Matlab. A 2D array MatriXX Evolution ionization chamber was used for dose verification. The 2D dose distributions were analyzed using OmniPro-I'mRT software and compared using the gamma index criteria of 3%/3mm and 3%/2mm with a low dose threshold of 10%. The correlation test results of complexity metrics and GPR show that the EAM correlates well with GPR for both cases. The correlation coefficient values for CNS cases are –0.918 (3%/3mm) and –0.864 (3%/2mm), while for breast cases the correlation coefficient values are –0.983 (3%/3mm) and –0.961 (3%/2mm). The MCS, CAM, CPA, and MU/Gy have a weak correlation with the value of the GPR."
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2022
T-pdf
UI - Tesis Membership  Universitas Indonesia Library
cover
Yahya Mustofa
"IMRT merupakan salah satu teknik radio terapi menggunakan pesawat linear akselerator dengan banyak lapangan penyinaran yang menggunakan intensitsradiasi yang berbeda-beda untuk mendapatkan dosis maksimal pada organ target tumor dan dosis seminimal mungkin pada organ sehat. Sebelum dilakukan penyinaran ke pasien, diperlukan verifikasi penyinaran IMRT antara perhitungan pada TPS dan pada keadaan sebenarnya di lapangan. Verifikasi dilakukan dengan menggunakan MatriXXEvolution. Dari penelitian yang telah dilakukan pada 5 pasien dengan klinis Glioblastoma Multiforme, dimana 3 pasien dilakukan verifikasi pada setiap lapangan dan gabungan semua lapangan, 2 pasien dilakukan verifikasi gabungan semua lapangan. Didapatkan kesesuaian piksel bagus untuk semua pasien dengan kriteria γ ≤1pada2%deltadose,dan2mmDTA.
Hasil verifikasi untuk semua lapangan penyinaran didapat kesesuaian piksel yaitu 99,64%; 99,81%; 99,69%; 99,35%; 99,67% untuk pasien 1-5. Kesesuaian piksel untuk verifikasi setiap lapangan penyinaran pada pasien 1; 99,38%; 95,47%; 99,63%; 98,98%; 99,86%; pasien 2; 96,56%; 98,65%; 99,54%; 99,63%; 98,00%; pasien 3; 98,67%; 97,70%; 99,81%; 99,52%; 99,26%. Perbedaan pengukuran disebabkan antara lain karena high dose gradient, daerah dosis rendah, dan penumbra. Pengukuran menggunakan detektor selain memiliki keuntungan dari waktu pengukuran yang lebih pendek, juga memiliki kelemahan ukuran detektor berdiameter 4,5mm dan jarak antara detektor7mm. Yang akan menunjukkan perbedaan besar ketika detektor melakukan pengukuran pada posisi tertentu seperti daerah dari dosis gradien, dosis rendah, penumbra sebagai detektor 4,5mm diameter."
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2011
S1041
UI - Skripsi Open  Universitas Indonesia Library
cover
Arif Budiman
"Telah dilakukan pengembangan sistem registrasi 2D/2D untuk verifikasi posisi pasien radioterapi berbasis bahasa C++ dengan tiga metode yaitu rigid transform, Viola-Wells mutual information dan Mattes mutual information. Implementasi sistem registrasi citra otomatis yang dikembangkan menggunakan data sekunder 34 pasien kanker nasofaring didapatkan hasil koreksi maksimum untuk rigid transform sebesar (2,79 ± 0,41) mm, Viola-Wells mutual information sebesar (2,77 ± 0,64) mm dan Mattes mutual information sebesar (2,69 ± 0,68) mm. Namun dalam proses uji validasi menggunakan fantom CIRS tidak berhasil dilakukan karena kontras citra EPID setelah melalui tahapan pre-processing tidak jauh berbeda.

Has been developed registration system 2D/2D for verifying patient’s radiotherapy position based on C++ languages which consist of 3 registration methods such as rigid transform, Viola-Wells mutual information and Mattes mutual information. Implementation of automatic image registration system that has been developed using secondary data from 34 Nasopharyngeal Cancer patients are obtained maximum correction at about (2,79 ± 0,41) mm, (2,77 ± 0,64) mm and (2,69 ± 0,68) mm for rigid transform, Viola-Wells mutual information and Mattes mutual information, respectively. However in the validation test using CIRS phantom is not success because the image contrast of EPID after pre-processing was not significantly different.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2014
S58654
UI - Skripsi Membership  Universitas Indonesia Library
cover
Mukhlisin
"Gerak tumor akibat pernapasan pasien merupakan masalah yang signifikan dalam pengobatan radioterapi kanker paru-paru, khususnya teknik radioterapi modern Intensity Modulated Radiotherapy (IMRT) dan Volumetric Modulated Arc Therapy (VMAT). Interaksi gerakan antara gerak target tumor dan MLC (interplay effect) memiliki keterbatasan dalam hal modulasi intensitas radiasi, probabilitasnya hanya sebagian kecil Planning Target Volume (PTV) menerima dosis radiasi sesuai perencanaan dosis Treatment Planning System (TPS) pada waktu tertentu.
Penelitian ini melakukan verifikasi dosimetri antara dosis yang direncanakan TPS dan dosis yang diterima volume tumor, akibat adanya interplay effect pada teknik IMRT dan VMAT. Penelitian menggunakan fantom toraks dinamik in-house dengan target tumor bergerak translasi arah superior-inferior dengan variasi amplitudo dan periode gerak tumor sebesar 9,3 mm dan 2,3 sekon, 20 mm dan 3,44 sekon, 30 mm dan 4,22 sekon. Pengukuran dosis titik dengan meletakkan dosimeter TLD-100 LiF:Mg,Ti dan Film Gafchromic EBT2 pada titik tengah target tumor dan organ at risk spinal cord. Penyinaran teknik IMRT menggunakan 7-field dan teknik VMAT menggunakan Rapidarc partial double arc dengan dosis preskripsi (95%) sebesar 200 cGy per fraksi.
Hasil penelitian menunjukkan bahwa kehadiran gerak target tumor paru-paru menyebabkan efek dosimetri yang tidak diinginkan berupa underdosage dalam volume tumor. Deviasi dosis rata-rata pada target tumor antara perencanaan dosis TPS dan hasil pengukuran pada teknik IMRT dengan target tumor bergerak statis, bergerak 9,3 mm, bergerak 20 mm, bergerak 30 mm berturut-turut sebesar 0,3% sampai 0,5%, -2,7% sampai -3,0%, -3,7% sampai -4,6%, dan -6,0% sampai -6,6%, sedangkan deviasi dosis pada teknik VMAT berturut-turut sebesar 0,2% sampai 0,9%, -1,6% sampai -1,9%, -2,9% sampai -3,1%, dan -5,0% sampai -5,3%. Hal berbeda, deviasi dosis untuk organ at risk spinal cord pada teknik IMRT berturutturut sebesar -5,6% sampai -1,0%, -6,8% sampai -6,9%, -3,7% sampai -5,9%, dan 0,7% sampai 1,0%, sedangkan deviasi dosis pada teknik VMAT berturut-turut sebesar -1,4% sampai -3,1%, -3,0% sampai -6,3%, -1,6% sampai -4,2%, dan 0,1% sampai 0,9%. Kenaikan amplitudo gerak target tumor menyebabkan dosis yang diterima volume tumor menurun. Namun sebaliknya, adanya kenaikan amplitudo gerak target tumor menyebabkan dosis yang diterima organ at risk spinal cord meningkat.

Tumor motion due to patient's respiratory is a significant problem in radiotherapy treatment of lung cancer, especially in modern radiotherapy treatment of Intensity Modulated Radiotherapy (IMRT) and Volumetric Modulated Arc Therapy (VMAT). The interplay effect is the effect that may occur as the motion of Linac (primarily the MLC) and motion of the tumor target interferes. At delivery dose treatment, a small part of Planning Target Volume (PTV) does not recover dose according to Treatment Planning System (TPS) Prescription.
This investigation was carried out through dosimetry verification between TPS and actual dose by tumor volume due to the interplay effect in IMRT and VMAT treatment. Tumor target of in-house dynamic thorax phantom was designed in linier sinusoidal motion toward superior-inferior direction with amplitude and period variation of tumor motion of 9,3 mm and 2,3s, 20 mm and 3,44s, 30 mm and 4,22s respectively. For point dose measurement, TLD-100 LiF:Mg,Ti and gafchromic EBT2 film detectors were took placed at midpoint of tumor target and spinal cord. IMRT treatment irradiation was applied by 7-fields and VMAT treatment by partial double arc, with prescription dose (95%) of 200 cGy per fraction.
The results showed that the occurrence of lung tumor target motion causes underdosage dosimetry effect in tumor volume. Mean dose deviation of tumor target between TPS and measurement in IMRT treatment by tumor target moves at condition of static, 9,3 mm, 20 mm and 30 mm were 0,3% to 0,5%, -2,7% to -3,0%, -3,7% to -4,6%, and -6,0% to -6,6% respectively while dose deviation in VMAT treatment were 0,2% to 0,9%, -1,6% to -1,9%, -2,9% to -3,1%, and -5,0% to -5,3% respectively. On the other hand, mean dose deviation of spinal cord in IMRT treatment were -5,6% to -1,0%, -6,8% to -6,9%, -3,7% to -5,9%, and 0,7% to 1,0% respectively and in VMAT treatment were -1,4% to -3,1%, -3,0% to -6,3%, -1,6% to -4,2%, and 0,1% to 0,9% respectively. The increment amplitude of tumor target motion reduced dose received by tumor volume and conversely, increased dose received by spinal cord.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2015
T43728
UI - Tesis Membership  Universitas Indonesia Library
cover
Mochamad Robby Fairuzzihab Qodarul
"Intensity modulated radiotherapy (IMRT) dan volumetric modulated arc therapy (VMAT) merupakan teknik radioterapi yang sering digunakan karena kompleksitasnya yang tinggi memungkinkan pemberian dosis yang maksimal pada target dengan meminimalkan dampak terhadap jaringan sekitar. Prosedur patient-specific quality assurance (PSQA) dibutuhkan untuk memastikan kesesuaian yang baik antara dosis perencanaan pada TPS dan dosis yang diterima saat pemberian pengobatan. PRIMO adalah perangkat lunak yang dapat melakukan PSQA dengan simulasi rekonstruksi berkas dynalog. Penelitian ini bertujuan untuk melakukan verifikasi dosis perencanaan awal dan hasil simulasi untuk memastikan dosis yang diberikan sesuai dengan yang direncanakan, serta mengevaluasi kriteria penerimaan yang telah ditetapkan sebelumnya. Sebanyak 5 data perencanaan dan berkas dynalog IMRT dan VMAT pasien kanker head and neck disimulasikan menggunakan PRIMO untuk mendapatkan distribusi dosis, kemudian dibandingkan dengan hasil perencanaan TPS untuk mendapatkan nilai gamma pass rate (GPR), percentage of agreement (PA) dan root-mean-square error (RMS) sehingga dapat dievaluasi. Perbandingan menunjukkan bahwa GPR 3%/3 mm dan 2%/2 mm memiliki hasil yang baik, dengan nilai rata-rata lebih besar dari 95% dan standar deviasi yang cukup kecil. Namun, evaluasi PA yang dilakukan menunjukkan hasil yang kurang memuaskan, banyak perbandingan struktur pada DVH yang menunjukkan nilai kurang dari 99% yang menjadi batas kelulusan verifikasi. Sedangkan RMS hampir tidak memiliki dampak langsung terhadap penyimpangan distribusi dosis yang dihasilkan dan memiliki nilai yang jauh lebih baik dari batas yang ditentukan. Metode verifikasi dosis menggunakan GPR dan RMS masih menjadi relevan karena mendapatkan hasil yang sesuai dengan rekomendasi, bahkan dapat menerapkan batas yang lebih ketat menggunakan kriteria GPR 2%/2 mm ≥95% dan RMS <1 mm. PA masih belum dapat menjadi evaluator tunggal, namun bisa menjadi pendukung dari parameter lainnya.

Intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) allow the delivery of maximum possible dose to the target while minimizing the influence on surrounding tissues. Patient-specific quality assurance (PSQA) is required to ensure good agreement between the planning dose by TPS and the dose received during treatment delivery. PRIMO can perform PSQA using dynalog files reconstruction simulation. This study aims to verify the simulation dose to ensure the dose delivered is in accordance with the planning dose, and to evaluate the previously established acceptance criteria. A total of 5 planning data and dynalog files for IMRT and VMAT head and neck cancer patients were simulated in PRIMO to obtain dose distribution, their results were compared to TPS planning results to obtain gamma pass rate (GPR), percentage of agreement (PA), and root-mean-square error (RMS) for evaluation. The comparison showed the 3%/3 mm and 2%/2 mm GPRs had good results, with average values greater than 95% and relatively narrow standard deviations. PA evaluation showed unsatisfactory results, with many comparisons on dose-volume histogram (DVH) having values less than 99% as the verification pass limit. Root-mean-square error (RMS) affected almost nothing in the deviation of simulated dose distribution and had better values than the specified limit. The dose verification method using GPRs and RMS was relevant because the evaluation results produced within the recommendations, stricter limits of GPR 2%/2 mm ≥95% and RMS <1 mm could be applied. The PA could not be the sole evaluator but complement other parameters for verification."
Depok: Fakultas Matematika Dan Ilmu Pengetahuan Alam Universitas Indonesia, 2024
T-pdf
UI - Tesis Membership  Universitas Indonesia Library