Mukhlisin
"Gerak tumor akibat pernapasan pasien merupakan masalah yang signifikan dalam pengobatan radioterapi kanker paru-paru, khususnya teknik radioterapi modern Intensity Modulated Radiotherapy (IMRT) dan Volumetric Modulated Arc Therapy (VMAT). Interaksi gerakan antara gerak target tumor dan MLC (interplay effect) memiliki keterbatasan dalam hal modulasi intensitas radiasi, probabilitasnya hanya sebagian kecil Planning Target Volume (PTV) menerima dosis radiasi sesuai perencanaan dosis Treatment Planning System (TPS) pada waktu tertentu.
Penelitian ini melakukan verifikasi dosimetri antara dosis yang direncanakan TPS dan dosis yang diterima volume tumor, akibat adanya interplay effect pada teknik IMRT dan VMAT. Penelitian menggunakan fantom toraks dinamik in-house dengan target tumor bergerak translasi arah superior-inferior dengan variasi amplitudo dan periode gerak tumor sebesar 9,3 mm dan 2,3 sekon, 20 mm dan 3,44 sekon, 30 mm dan 4,22 sekon. Pengukuran dosis titik dengan meletakkan dosimeter TLD-100 LiF:Mg,Ti dan Film Gafchromic EBT2 pada titik tengah target tumor dan organ at risk spinal cord. Penyinaran teknik IMRT menggunakan 7-field dan teknik VMAT menggunakan Rapidarc partial double arc dengan dosis preskripsi (95%) sebesar 200 cGy per fraksi.
Hasil penelitian menunjukkan bahwa kehadiran gerak target tumor paru-paru menyebabkan efek dosimetri yang tidak diinginkan berupa underdosage dalam volume tumor. Deviasi dosis rata-rata pada target tumor antara perencanaan dosis TPS dan hasil pengukuran pada teknik IMRT dengan target tumor bergerak statis, bergerak 9,3 mm, bergerak 20 mm, bergerak 30 mm berturut-turut sebesar 0,3% sampai 0,5%, -2,7% sampai -3,0%, -3,7% sampai -4,6%, dan -6,0% sampai -6,6%, sedangkan deviasi dosis pada teknik VMAT berturut-turut sebesar 0,2% sampai 0,9%, -1,6% sampai -1,9%, -2,9% sampai -3,1%, dan -5,0% sampai -5,3%. Hal berbeda, deviasi dosis untuk organ at risk spinal cord pada teknik IMRT berturutturut sebesar -5,6% sampai -1,0%, -6,8% sampai -6,9%, -3,7% sampai -5,9%, dan 0,7% sampai 1,0%, sedangkan deviasi dosis pada teknik VMAT berturut-turut sebesar -1,4% sampai -3,1%, -3,0% sampai -6,3%, -1,6% sampai -4,2%, dan 0,1% sampai 0,9%. Kenaikan amplitudo gerak target tumor menyebabkan dosis yang diterima volume tumor menurun. Namun sebaliknya, adanya kenaikan amplitudo gerak target tumor menyebabkan dosis yang diterima organ at risk spinal cord meningkat.

Tumor motion due to patient's respiratory is a significant problem in radiotherapy treatment of lung cancer, especially in modern radiotherapy treatment of Intensity Modulated Radiotherapy (IMRT) and Volumetric Modulated Arc Therapy (VMAT). The interplay effect is the effect that may occur as the motion of Linac (primarily the MLC) and motion of the tumor target interferes. At delivery dose treatment, a small part of Planning Target Volume (PTV) does not recover dose according to Treatment Planning System (TPS) Prescription.
This investigation was carried out through dosimetry verification between TPS and actual dose by tumor volume due to the interplay effect in IMRT and VMAT treatment. Tumor target of in-house dynamic thorax phantom was designed in linier sinusoidal motion toward superior-inferior direction with amplitude and period variation of tumor motion of 9,3 mm and 2,3s, 20 mm and 3,44s, 30 mm and 4,22s respectively. For point dose measurement, TLD-100 LiF:Mg,Ti and gafchromic EBT2 film detectors were took placed at midpoint of tumor target and spinal cord. IMRT treatment irradiation was applied by 7-fields and VMAT treatment by partial double arc, with prescription dose (95%) of 200 cGy per fraction.
The results showed that the occurrence of lung tumor target motion causes underdosage dosimetry effect in tumor volume. Mean dose deviation of tumor target between TPS and measurement in IMRT treatment by tumor target moves at condition of static, 9,3 mm, 20 mm and 30 mm were 0,3% to 0,5%, -2,7% to -3,0%, -3,7% to -4,6%, and -6,0% to -6,6% respectively while dose deviation in VMAT treatment were 0,2% to 0,9%, -1,6% to -1,9%, -2,9% to -3,1%, and -5,0% to -5,3% respectively. On the other hand, mean dose deviation of spinal cord in IMRT treatment were -5,6% to -1,0%, -6,8% to -6,9%, -3,7% to -5,9%, and 0,7% to 1,0% respectively and in VMAT treatment were -1,4% to -3,1%, -3,0% to -6,3%, -1,6% to -4,2%, and 0,1% to 0,9% respectively. The increment amplitude of tumor target motion reduced dose received by tumor volume and conversely, increased dose received by spinal cord.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2015
T43728
UI - Tesis Membership  Universitas Indonesia Library
cover
Aloysius Mario Yudi Putranto
"Penelitian ini bertujuan memeriksa efek dosimetris dari pergerakan tumor untuk beberapa variasi besar ampiltudo pergerakan. Penelitian ini menggunakan fantom dinamik toraks yang dibangun berdasarkan CIRS Dynamic Thorax Phantom model 008A. Fantom ini mensimulasikan pergerakan secara translasi pada arah Superior dan Inferior (SI) serta rotasi pada bidang Lateral kanan-kiri (RL) dan Anterior Posterior (AP) untuk menirukan pergerakan akibat pernafasan manusia. Fantom ini didesain dan dikontrol menggunakan  linear actuator, servo motor, Adafruit motor shield L293D dan Arduino UNO R3. Fantom ini digunaan untuk mengevaluasi dosis titik menggunakan teknik 3D-CRT, IMRT, VMAT dan amplitudo translasi 5 mm; 10 mm; 15 mm  dengan amplitudo rotasi 90°. Pengukuran dosis titik menggunakan dosimeter film GafChromic EBT 3 dan thermoluminescence TLD LiF-100. Fantom toraks dinamik yang dibuat dapat meniru pergerakan translasi dengan amplitudo 5±0,2 mm; 10±0,5 mm; 15±0,4 mm dan pergerakan rotasi dengan amplitudo 89°±2°. Dari penelitian ini didapatkan penggunaan Internal Margin sebesar 5 mm mampu memberikan kompensasi deviasi dosis berkisar antara 0,6% sampai dengan 1%, penurunan dosis terbesar adalah pada arah Superior dan Inferior, gerak rotasi target tumor berpengaruh pada peningkatan deviasi dosis pada lateral kanan dan kiri akibat pebedaan kedalaman serta Teknik VMAT tidak disarankan untuk target bergerak karena memberikan deviasi terbesar yaitu berkisar pada -5% sampai dengan 9% akibat dari derajat kebebasan pergerakan antara gantry dan target semakin tinggi.

This study aimed to investigate dosimetric effect of target movement for several translational amplitude. This work used in-house dynamic thorax phantom based on CIRS Dynamic Thorax Phantom model 008A. This phantom simulated translation in superior-inferior direction, rotational in the anterior-posterior and left-right lateral plane to mimic human respiratory motion. It was designed and controlled by linear actuator motor, servo motor, Adafruit motor shield L293D and Arduino UNO R3. It was implemented to evaluate point dose of 3D-CRT, IMRT, VMAT technique and for 5 mm; 10 mm; 15 mm translational motion amplitude 90° rotational target motion amplitude. The point dose measurement used GafChromic EBT 3 film and TLD LiF-100 as dosimeter. This in-house dynamic thorax phantom can mimicking NSCLC target movement for translational amplitude 5±0,2 mm; 10±0,5 mm; 15±0,4 mm and rotational amplitude 89°±2° From this study it found that the use of Internal Margin of 5 mm was able to compensate dose deviation about 0.6% to 1%, the largest decrease in dosage was in the Superior and Inferior direction, the rotational motion of the target tumor had an effect on increasing dose deviation on the right and left lateral due to the difference in depth and VMAT Technique it is not recommended for moving targets because it provides the largest deviation which ranges from -5% to 9% due to the higher degree of freedom of movement between the gantry and the target."
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2019
T53015
UI - Tesis Membership  Universitas Indonesia Library
cover
Adi Teguh Purnomo
"Evaluasi dosis radiasi yang diterima pasien dapat dilakukan menggunakan metode dosimetri in-vivo dengan melakukan pengukuran dosis masukan yang didefinisikan sebagai dosis serap pada kedalaman maksimum. Dosimeter in-vivo yang digunakan adalah TLD-100 rod, Film Gafchromic EBT2, dan dioda EDP-153G yang dikalibrasi berdasarkan protokol IAEA dan ESTRO. Hasil faktor kalibrasidetektor dioda sebesar 0.0988 cGy/ADU dan 0.0966 cGy/ADU untuk protokol IAEA dan ESTRO sedangkan Faktor kalibrasidetektor TLD memiliki rata-rata perbedaan sebesar 3.69 ± 0.04 %. Selain itu, faktor koreksi dioda EDP-153G terhadap variasi SSD, luas lapangan, dan linieritas dosis berada dalam rentang nilai 0.99-1.01 untuk protokol IAEA dan ESTRO sedangkan faktor koreksi terhadap sudut sinar datang berada dalam rentang nilai 1.028 - 1.037 dan 1.027-1.057 untuk protokol IAEA dan ESTRO. Pengukuran dosimetri in-vivo dititikberatkan pada kasus kanker paru dan prostat dengan meletakkan dosimeter in-vivo diatas permukaan kulit untuk beberapa titik pengukuran. Dosis masukan yang diperoleh bervariasi untuk setiap titik pengukuran akibat adanya perbedaan fluence. Evaluasi dosis pada organ target juga ditentukan dengan meletakkan film Gafchromic EBT2 pada slab fantom Rando Alderson. Persentase error yang diperoleh terhadap dosis yang direncanakan yakni sebesar 0.03% pada film 1 dan 2.5% pada film 2 untuk kanker paru sedangkan untuk kanker prostat sebesar 5.88% untuk film 1 dan 5.50% untuk film 2.

Evaluation of the radiation dose received by the patient could be performed with the in-vivo dosimetry by mean of the entrance dose measurement which defined as the absorbed dose to the maximum depth. TLD-100 rod, Gafchromic EBT2 film, and diode EDP-153G were used as in-vivo dosimeters and calibrated by the IAEA and ESTRO protocol. The results of calibration factor for diode EDP-153G was 0.0988 cGy/ADU and 0.0966 cGy/ADU for the IAEA and ESTRO, respectively while the differences of mean percentage was 3.69 ± 0.04 % relative to IAEA protocol for TLD. In addition, a correction factor for diode EDP-15­3Gto the variation of SSD, field size, and dose linearity were within the range of0.99-1.01 for the IAEA and ESTRO while the angular correction factor was at 1.028-1.037 and 1.027-1.057 for the IAEA and ESTRO protocol, respectively. The in-vivo dosimetry measurement was concerned in lung and prostate carcinoma by putting the in-vivo dosimeters on skin surface at some points of interest. The entrance dose measurement was varied for each point of measurement due to difference of fluence. The target dose evaluation was also determined by placing the Gafchromic EBT2 Film into slab of Rando Alderson phantom. The percentage error to the planned dose was 0.03% at the film 1 and 2.5%at the film 2 for lung carcinoma while for prostate carcinoma was 5.88% for the film 1 and 5.50% for film 2.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2014
S54915
UI - Skripsi Membership  Universitas Indonesia Library
cover
Pamungkas Hudigomo
"Telah dilakukan penelitian yang bertujuan untuk mengimplementasikan film Gafchromic XR-RV3 untuk verifikasi dosis radioterapi pada IMRT dan VMAT. Evaluasi dosis pada target ditentukan dengan meletakkan film Gafchromic XRRV3 dan EBT2 padqa slab fantom Rando Alderson. Percobaan dilakukan dengan menggunakan Varian Clinac Trilogy, Inc. Proses simulasi terlebih dahulu dilakukan pada fantom Rando Alderson bagian thorax dan pelvis di CT simulator Phillips, Inc. Dua kasus perencanaan IMRT dan VMAT dibuat menggunakan TPS Eclipse ver. 10. Film Gafchromic yang telah dipapar kemudian dipindai menggunakan scanner Epson Perfection V700 ke dalam format tagged image file (.TIFF) dengan 72 dpi dan RGB 48 bit yang kemudian dianalisis menggunakan algoritma in-house yang telah dikembangkan oleh penelitian sebelumnya. Dalam penelitian ini dilakukan perbandingan pengukuran distribusi dosis pada film Gafchromic XR-RV3 dan EBT2 untuk mendapatkan dosis rata-rata dalam bentuk kurva histogram. Hasil penelitian ini memiliki persentase kesalahan terhadap dosis preskripsi pada kasus kanker prostat sebesar -4.85% pada film Gafchromic EBT2 dan -1.94% pada film Gafchromic XR-RV3 dengan teknik penyinaran IMRT, sedangkan untuk teknik penyinaran VMAT sebesar -4.48% pada film Gafchromic EBT2 dan -7.47% pada film Gafchromic XR-RV3. Di sisi lain, persentase kesalahan terhadap dosis preskripsi pada kasus kanker paru-paru dengan teknik penyinaran IMRT adalah 14.47% pada film Gafchromic EBT2 dan -4.37% pada film Gafchromic XR-RV3, sedangkan untuk teknik penyinaran VMAT sebesar 51.64% pada film Gafchromic EBT2 dan -28.07% pada film Gafchromic XR-RV3.

The research aims to implement Gafchromic XR-RV3 films in order to verify the dose of radiotherapy on IMRT and VMAT. The evaluation of targeted dose is determined by putting films of Gafchromic XR-RV3 and EBT2 on Rando Alderson slab fantom. The experiment was performed with Varian Clinac Trilogy, Inc. The simulation process was initially conducted on thorax and pelvis on Phillips Inc.’s CT Simulator. Two cases of IMRT and VMAT plans were made using Eclipse TPS ver. 10. Exposed Gafchromic films then scanned using Epson Perfection V700 into (.TIF) format in 72 dpi and RGB 48 bit which was analyzed by in-house algorithm that had been developed on a previous research. The comparison between EBT2 and XR-RV3 Gafchromic films were used to obtain average dose in the form of histogram curve. The percentage of errors in the case of prostate cancer toward the planned dose were -4.85% in EBT2 Gafchromic films and -1.94% in the XR-RV3 Gafchromic films on IMRT technique, whereas for VMAT were -4.48% on EBT2 Gafchromic films and -7.47% on XR-RV3 Gafchromic films. On the other hand, the percentage of errors in the case of lung cancer toward the planned dose with IMRT technique were 14.47% on EBT2 Gafchromic films and -4.37% on XR-RV3 Gafchromic films, while for VMAT were 51.64% on EBT2 Gafchromic films and -28.07% on XR-RV3 Gafchromic films.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2014
S57739
UI - Skripsi Membership  Universitas Indonesia Library
cover
Ade Abdul Wahab
"Pergerakan respirasi saat treatment radioterapi tumor dalam toraks seperti pada kasus Non-small Cell Lung Cancer (NSCLC) merupakan salah satu hal yang perlu dipertimbangkan dalam keakurasian dosis. Pergerakan tersebut menghasilkan ketidakpastian posisi target pada saat treatment radiasi yang dapat mengakibatkan adanya kekurangan dosis akibat adanya dose blurring dan interplay effect. Penelitian terdahulu membuktikan bahwa penambahan margin internal pada PTV dapat mengurangi perbedaan dosis perencanaan dengan dosis terukur. Penelitian ini bertujuan untuk mempelajari pengaruh metode optimasi robust yang memperhitungkan ketidakpastian posisi relatif terhadap volume target dalam perencanaan radioterapi dan perbandingannya dengan metode konvensional yang dilakukan pada fantom toraks dinamik in-house. Penggunaan metode ini meningkatkan nilai konformitas dosis pada teknik IMRT dan VMAT, meningkatkan uniformitas dosis pada teknik VMAT, namun mengurangi kecuraman dosis fall-off perencanaan. Metode ini meningkatkan nilai dosis target pada perencanaan IMRT konvensional dari rerata 199,63 cGy menjadi 200,43; 206,26; dan 204,20 cGy untuk ketidakpastian 5 mm, 10 mm, dan 15 mm, tetapi menurunkan dosis target pada perencanaan VMAT konvensional dari rerata 201,59 cGy menjadi 198,84; 199,05; dan 199,06 cGy untuk ketidakpastian 5 mm, 10 mm, dan 15 mm. Metode ini juga meningkatkan dosis OAR di area terdekat target dan menurunkan dosis di area lainnya pada teknik IMRT, namun hampir seluruh area OAR mendapatkan peningkatan dosis pada teknik VMAT.

Respiration movement, when treating radiotherapy to the tumor in the thorax, such in the case of NSCLC, is one of the challenges of dose accuracy. The target movement during radiation treatment causes dose blurring and interplay effects. A previous study proved that internal margins on PTV reduce the differences between plan and measured doses. This study aims to observe the effect of robust optimization, which provide uncertainty of target volume position on photon planning. We compared the robust with a conventional method on the in-house thorax dynamic phantom. This method enhanced the dose conformity index in the IMRT and VMAT techniques and the dose uniformity in the VMAT technique. However, robust planning reduces the steepness of the dose in both IMRT and VMAT. The average target dose of IMRT technique was increased from 199.63 cGy in conventional planning to 200.43; 206.26; and 204.20 cGy for planning with an uncertainty of 5 mm, 10 mm and 15 mm. On the other hand, the average target dose was reduced from 201.59 cGy in conventional planning to 198.84; 199.05; and 199.06 cGy with an uncertainty of 5 mm, 10 mm and 15 mm for VMAT techniques, respectively. We found that the robust plan delivered a higher dose at OAR near the target than the conventional plan, but the dose is lower in the other areas on the IMRT technique. For VMAT, most of the OAR achieved a higher dose on a robust photon plan."
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2020
S-pdf
UI - Skripsi Membership  Universitas Indonesia Library
cover
Aditya Fergiawan
"Penelitian ini bertujuan untuk mempelajari karakteristik dosimetri dari film gafchromic XR-RV3 yang memiliki sensitifitas terhadap rentang dosis 1 cGy hingga 30 Gy serta membandingkannya dengan film EBT2 dengan rentang dosis 1 cGy hingga 40 Gy untuk verifikasi dosis pada Radioterapi. Karakteristik film seperti respon terhadap dosis tinggi, pengaruh energi serta orientasi film menjadi fokus penelitian. Pengukuran dilakukan menggunakan energi Megavoltage x-rays (6 MV dan 10 MV), dan Cobalt 60 yang sudah dikalibrasi sesuai dengan protokol dosimetri IAEA TRS-398. Respon film pada masing-masing energi diukur pada rentang dosis 50 cGy hingga 10 Gy. Pemindaian pada masing-masing film menggunakan EPSON Perfection V700 flatbed scanner, mode transmisi untuk film EBT2 dan mode refleksi untuk film XR-RV3, 48 bit color, dan resolusi spasial 75 dpi. Data yang diperoleh dianalisa menggunakan software FilmQA Pro, ROI berukuran 3 x 3 cm2 pada bagian tengah dari film untuk memperoleh nilai pixel rata-rata. Untuk masing-masing nilai pixel yang diperoleh dari masing-masing film selanjutnya ditentukan nilai rata-rata net Optical Density (netOD). Respon bacaan film EBT2 dan XR-RV3 pada rentang energi yang berbeda menunjukkan pengaruh energi yang rendah dengan deviasi rata – rata sebesar ±1%. Hasil perbandingan karakteristik respon bacaan film EBT2 dan XR-RV3 untuk masing-masing modalitas energi berbeda menunjukkan deviasi sebesar ±3%. Pengujian respon bacaan film XR-RV3 pada sisi orientasi yang berbeda tidak menunjukkan perbedaan yang signifikan dengan deviasi rata – rata sebesar 0.2%. Hasil evaluasi profil berkas menggunakan film XR-RV3 menunjukkan jumlah noise (disturbance) yang lebih banyak dibandingkan dengan EBT2. Pengukuran nilai simetris dan kerataan berkas pada rentang energi Cobalt-60 menunjukkan penyimpangan yang cukup besar pada posisi film horizontal terhadap sumbu utama berkas yaitu diatas 30%. Pengujian sensitifitas respon bacaan film XR-RV3 pada dosis tinggi dan dosis rendah menunjukkan hasil yang lebih baik dibandingkan dengan film EBT2.

The purpose of this study is to compare the dosimetry characterization of XR-RV3 film with its sensitivity to dose levels from 1 cGy to 30 Gy and EBT2 film with its sensitivity to dose levels from 1 cGy to 40 Gy for dose verification in radiation therapy. The characteristics of film such as response at high dose levels, energy dependence and side orientation dependence to scanner were the main focus of this study. The film response to each energy (foton 6 MV, 10 MV and Cobalt-60) was measured over the dose levels from 50 cGy to 10 Gy and the output source were calibrated following IAEA TRS-398 Dosimetry Protocol. Each film piece was scanned using EPSON Perfection V700 flatbed scanner, 48-bit color, 75 dpi spatial resolution, Transmission mode for EBT2 film and Reflection mode for XR-RV3 film. The data were analyzed using FilmQA Pro Software and for each scanned image, a ROI size 3 x 3 cm2 at the field center was selected to obtain the mean pixel value. Furthermore, the pixel value from each film were calculated to find the average of net Optical Density (netOD). From this study, the result revealed that there was no significant difference in characteristics response between XR-RV3 and EBT2 film with standar deviation of ±3%. The energy dependence of XR-RV3 and EBT2 film was found to be relatively small within measurement uncertainties ±1%. The dependence of XR-RV3 film side orientation is negligibly small with the standard deviation of 0.2%. The result of beam profile dosimetry evaluation using XR-RV3 showed more amount of noise or disturbance than EBT2 film. The measurement of beam symmetry and beam flatness on energy range of Cobalt-60, showed a significant error of film in horizontal position at central axis beam that was above 30%. In addition, the sensitivity of response XR-RV3 film at high and low dose shown better result than EBT2 film.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2014
S57939
UI - Skripsi Membership  Universitas Indonesia Library
cover
Febrianti M
"ABSTRAK
Penelitian ini bertujuan untuk memverifikasi distribusi dosis pada teknik 3D-CRT dan IMRT dengan menggunakan dosimeter TLD rod dan film gafchromic EBT3 pada kasus payudara yang telah dibuat di TPS berdasarkan data klinis di Siloam Hospital TB Simatupang. Penelitian ini dilakukan menggunakan Linac Trilogy dengan TPS menggunakan modul Eclipse versi 11 yang memproduksi berkas foton dengan radiasi 6 dan 10 MV. Film EBT3 dengan ukuran 4 4 cm2 diiradiasi menggunakan energi 6 MV dalam variasi dosis 0 - 260 cGy. Penelitian ini menggunakan lapangan radiasi 10 10 cm2. Film dipindai menggunakan flatbed scanner Epson V700 dalam evaluasi pixel value. Perhitungan dan plot antara netOD dan dosis untuk mendapatkan kurva kalibrasi yang digunakan untuk memperoleh dosis pengukuran pada pasien. TLD rod dengan ukuran 3 mm diiradiasi dengan energi 6 MV dalam dosis 200 cGy. TLD dibaca menggunakan TLD reader Harshaw dalam bentuk muatan coulomb. Perhitungan dan plot antara muatan dan dosis akan mendapatkan faktor kalibrasi yang digunakan untuk memperoleh dosis pengukuran pada pasien. TLD dan film EBT3 ditempelkan pada 5 titik di masker yang digunakan pasien dimana titik 1 daerah trakea, titik 2 daerah aksila, titik 3 daerah mammaria interna, titik 4 daerah nipple, dan titik 5 mammae inferior. Hasil perbedaan dosis menunjukkan penggunaan film gafchromic EBT3 pada verifikasi dosis titik kasus kanker payudara lebih baik daripada TLD rod pada teknik 3D-CRT dan teknik IMRT. Perbedaan dosis antara TPS dan film EBT3 dalam rentang 0,66 sampai dengan 5,67 teknik 3D-CRT dan -0,67 sampai dengan -7,57 teknik IMRT . Perbedaan dosis antara TPS dan TLD rod dalam rentang -2,63 sampai dengan 29,14 teknik 3D-CRT dan -16,41 sampai dengan -46,84 teknik IMRT . Uji statistika menunjukkan korelasi kuat 1,00 tetapi tidak signifikan sig < 0,05 serta terdapat perbedaan dosis titik rata-rata yang signifikan antara film EBT3 dan TLD rod pada kelompok pasien sebelum dan setelah dilakukan operasi. Tidak terdapat perbedaan dosis titik rata-rata antara kelompok sebelum dan setelah dilakukan operasi baik teknik 3D-CRT maupun IMRT dan tidak terdapat perbedaan dosis titik rata-rata antara teknik 3D-CRT dan IMRT.

ABSTRACT
The aim of thisresearch is to verify dose distribution in 3D CRT and IMRT by using TLD rod dosimeter and Gafchromic EBT3 films on breast cases that have been made in TPS based on clinical data at Siloam Hospital TB Simatupang.This research was conducted using Linac Trilogy with TPS using Eclipse module version 11 which produces photon beam with radiation of 6 and 10 MV.EBT3 films of size 4 4 cm2 were irradiated using 6MV photon beams in dose variation of 0 260 cGy. Radiation field of size 10 10 cm2 was used in this research. The irradiated EBT3 film was scanned using Epson V700 flatbed scanner to evaluate the pixel value. The calculation and plot between netOD and dose was used to get the calibration curve which will be used to obtain dose measurement for patients. TLD rod of size 3 mm was irradiated using 6 MV in dose variation of 200 cGy. TLD was read by using HarshawTLD reader in coulomb. The calculation and plot between netOD and dose will generate calibration factor used to get dose measurement for patients. TLD and EBT3 film were taped to 5 points in the mask used by patients where 1 point is the OAR trachea and 4 points is the PTV50. The results of dose difference at PTV50 with point orientation.TLD and EBT3 film were taped to 5 points in the mask used by patients wherethe first point is in the OAR trachea , the second point is in the axilla, the third point is in the mammaria interna, the fourth point is in the nipple, and the fifth point is in the mammae inferior. The result of differences in dose shows the use of gafchromic EBT3 film in dose verification on breast cancer is better than TLD rod in 3D CRT technique and IMRT technique. The dose difference between TPS and EBT3 film ranged from 0.66 to 5.67 3D CRT technique and 0.67 to 7.57 IMRT technique . The dose difference between TPS and TLD rod is in the range 2.63 to 29.14 3D CRT technique and 16.41 up to 46.84 IMRT technique . The statistical test showed a strong correlation 1.00 but insignificant sig "
2017
T47602
UI - Tesis Membership  Universitas Indonesia Library
cover
Aninda Fitriandini
"Telah dilakukan penelitian pengukuran dosis titik (point dose) pada daerah target (paru-paru), jantung, dan tulang belakang menggunakan empat dosimeter (PTW N30013, Exradin A16, TLD, dan film Gafchromic EBT2). Tujuan utama dari penelitian ini adalah membandingkan keempat sistem dosimeter tersebut dalam hal akurasi, presisi, rentang pengukuran, serta resolusi spasial dan ukuran fisik. Pengukuran dosis dilakukan hanya pada satu kali fraksi penyinaran. Hasil pengukuran menunjukkan bahwa TLD memiliki akurasi dan presisi yang paling rendah. Adapun efek volume averaging pada bilik ionisasi menyebabkan nilai diskrepansi mencapai -13.30% pada daerah target. Sedangkan film EBT2, dengan nilai diskrepansi <1% pada 3D-CRT dan IMRT, dinyatakan sebagai dosimeter alternatif yang sesuai untuk digunakan pada verifikasi dosis titik.
This research was conducted by measuring point dose in the target area (lungs), heart, and spine using four dosimeters (PTW N30013, Exradin A16, TLD, and the Gafchromic EBT2 film). The main objective of this study was to compare the dosimetry of those different systems. Dose measurements performed only in single fraction of irradiation. The measurement results shown that TLD has the least accuracy and precision. As the effect of the volume averaging, ionization chamber reaches the discrepancy value up to -13.30% in the target area. EBT2 film has discrepancy value of <1% in the 3D-CRT and IMRT techniques. This dosimeter is proposed to be an appropriate alternative dosimeter to be used at dose point verification."
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2015
S59860
UI - Skripsi Membership  Universitas Indonesia Library
cover
Retno Zurma
"[ABSTRAK
Film gafchromic EBT2 dan EBT3 cocok digunakan untuk jaminan kualitas (QA) pada treatment planning systems (TPS) dan linear accelerator (LINAC) serta verifikasi radioterapi teknik IMRT dan VMAT. Namun generasi selanjutnya dari film gafchromic EBT yaitu EBT2 dan EBT3 ketika dilakukan scanning (pemindaian) menggunakan scanner flatbed masih terdapat artefak geometris yang dahulunya juga ditemukan pada film EBT sehingga memerlukan penanggulangan dan koreksi yang sesuai. Pada penelitian ini menggunakan scanner flatbed Epson expression 10000XL, Epson perfection V700 dan Mikrotek 1000XL plus yang dapat menangkap warna dengan rinci dan presisi. Perangkat lunak yang digunakan dalam koreksi ini adalah FilmQA Pro 2015. Hasil yang diperoleh yaitu Koefisien A dan B diterapkan untuk variasi posisi lateral terhadap nilai tanggapan di lokasi yang bersamaan, sehingga pengkoreksian tanggapan artifak lateral dapat dilakukan. Pada dosis maksimum nilai koreksinya <0,5% dan pada kasus <0.2%. Pada pengukuran ulang untuk setiap film, standar deviasi yang diperoleh untuk setiap film sekitar 0.19%. Pengujian dosimetri channel tunggal (merah) sebelum dan sesudah koreksi pada kasus IMRT didapatkan hampir 91% dari piksel memenuhi 3% / 3mm kriteria gamma keseluruhan dengan ambang batas dosis > 10%. Untuk pengujian kasus VMAT, film di scan dengan posisi di pusat scanner dan posisi ektrim tepi scanner, maka pengukuran di channel warna merah 4% lebih tinggi dibandingkan channel warna hijau dan biru. Rata-rata konsistensi pengukuran dosis selama daerah terkena dosis > 100 cGy adalah sekitar 21,8 cGy dan di daerah yang paling dekat dengan tepi scan window sekitar 10 mm dengan perbedaan dosis 35 cGy. Setelah dilakukan koreksi, rata-rata konsistensi dosis pada tiga channel di sekitar daerah terpapar sekitar 5,1 cGy. Perbedaan dosis antar channel sekitar 9 cGy. Hasil ini menunjukkan koreksi respon artefak lateral memang perlu dilakukan. Respon yang diukur tergantung pada posisi film pada pemindai dan dosis yang diberikan. Pengukuran pada channel warna merah menunjukkan sensitivitas yang lebih besar pada dosis rendah, sedangkan respon pada channel hijau atau biru memberikan perpanjangan jangkauan dinamis dari film ke dosis tinggi. Metode triple-channel dosimetri telah terbukti memiliki keuntungan signifikan atas single channel dosimetry dengan akurasi dosimetri yang baik.

ABSTRACT
The film of gafchromic EBT2 and EBT3 are appropriate for quality assurance (QA) to the radiotherapy technics verification of IMRT and VMAT. In the next generations of EBT which are EBT2 and EBT3, are still attained the geometrical artifacts once scanning by flatbed scanner which is used to be found on EBT film in order to need an overcome ways and an appropriate correction. This research used a flatbed scanner EPSON expression 10000XL, EPSON perfection V700 and Microtech 1000XL plus and FilmQA Pro 2015 as a software for correcting it. The result shows that coefficient A and B were applied for lateral correction artifacts responses can be made. On the maximum dose, it has correction value as much as <0,5% and in case was <0.2%. The deviation standard was approximately 0,19% on re-measure for each film. Prior to the dosimetry measurement of single channel (red) and subsequent to the correction IMRT has got almost 91% of pixel which met with gamma criteria, 3% / 3mm, with dose threshold 10% TH. In the measurement of VMAT‟s case, film scanned on the center of scanner and the extreme position of the scanner edge. Thus, measurement in the red channel was 4% higher than green and blue channel. The average of dose measurement consistency during area exposed to doses > 100 cGy is approximately 21,8 cGy and the closest area with the edge of scan window is around 10 mm at difference doses 35 cGy. After correction, the average of doses consistency was 5,1 cGy on the three channel where exposed. The distinction of doses was approximately 9 cGy among channels. This result stated that the correction of artifacts lateral response is needed to carry on. The method of triple-channel dosimetry has the significant advantage on single-channel dosimetry with a good consistency. Through escalation dose and variation of lateral position, the distinct response would be increased on single-channel which is red channel. The measurement on red channel reveals the highest sensitivity on the slight dose. Whereas, a green or blue channel response provides a dynamic range extension of film to a high dose. Thus, the method of dosimetry triple-channel has the significant contribution of single-channel evidently with a good dosimetric accuracy., The film of gafchromic EBT2 and EBT3 are appropriate for quality assurance (QA) to the radiotherapy technics verification of IMRT and VMAT. In the next generations of EBT which are EBT2 and EBT3, are still attained the geometrical artifacts once scanning by flatbed scanner which is used to be found on EBT film in order to need an overcome ways and an appropriate correction. This research used a flatbed scanner EPSON expression 10000XL, EPSON perfection V700 and Microtech 1000XL plus and FilmQA Pro 2015 as a software for correcting it. The result shows that coefficient A and B were applied for lateral correction artifacts responses can be made. On the maximum dose, it has correction value as much as <0,5% and in case was <0.2%. The deviation standard was approximately 0,19% on re-measure for each film. Prior to the dosimetry measurement of single channel (red) and subsequent to the correction IMRT has got almost 91% of pixel which met with gamma criteria, 3% / 3mm, with dose threshold 10% TH. In the measurement of VMAT‟s case, film scanned on the center of scanner and the extreme position of the scanner edge. Thus, measurement in the red channel was 4% higher than green and blue channel. The average of dose measurement consistency during area exposed to doses > 100 cGy is approximately 21,8 cGy and the closest area with the edge of scan window is around 10 mm at difference doses 35 cGy. After correction, the average of doses consistency was 5,1 cGy on the three channel where exposed. The distinction of doses was approximately 9 cGy among channels. This result stated that the correction of artifacts lateral response is needed to carry on. The method of triple-channel dosimetry has the significant advantage on single-channel dosimetry with a good consistency. Through escalation dose and variation of lateral position, the distinct response would be increased on single-channel which is red channel. The measurement on red channel reveals the highest sensitivity on the slight dose. Whereas, a green or blue channel response provides a dynamic range extension of film to a high dose. Thus, the method of dosimetry triple-channel has the significant contribution of single-channel evidently with a good dosimetric accuracy.]"
2016
T45204
UI - Tesis Membership  Universitas Indonesia Library
cover
Ahmad Maulana
"[ABSTRAK
Telah dilakukan verifikasi dosis organ target dan jaringan sehat di sekitar target dengan menempatkan TLD Rod LiF100 dan film Gafchromic EBT2 di lubang slab bagian pelvis dari phantom Rando Alderson untuk simulasi kanker prostat. TLD dievaluasi menggunakan TLD Reader Harshaw, sementara Film Gafchromic EBT2 dipindai menggunakan scanner Epson Perfection V700 dengan mode transmisi, red channel dan resolusi 72 dpi. Pengukuran dosis titik dilakukan dengan membandingkan antara dosis yang direncanakan TPS Eclipse ver. 11 dan dosis yang diukur pada target organ target dan organ beresiko menggunakan teknik IMRT dan VMAT. Hasilnya adalah deviasi dosis pada organ target menggunakan teknik IMRT dan VMAT adalah kurang dari 5%. Demikian pula, deviasi dosis pada bladder dan rectum untuk kedua teknik juga kurang dari 5% karena posisinya sangat dekat dengan target volume. Di sisi lain, deviasi dosis di femoral head lebih dari 5% untuk kedua teknik karena lokasinya pada gradien dosis rendah. Selanjutnya, deviasi dosis organ target untuk teknik IMRT cenderung lebih kecil dari teknik VMAT baik untuk TLD dan Film. Perbedaan dosis pada dosis titik organ target antara IMRT dan VMAT kurang dari 1% tetapi terjadi pada dosis yang random untuk organ beresiko. Adapun dosis permukaan pada teknik IMRT cenderung lebih kecil dari teknik VMAT jika kita menggunakan TLD, tetapi dosis pada film EBT2 cenderung sama antara teknik IMRT dan VMAT.

ABSTRACT
Have been done the dose verification of the target and healthy tissues around by placing the TLD Rod LiF100 and EBT2 Gafchromic film at slab hole of pelvic part of the Alderson Rando phantom for prostate cancer simulation. The Exposed TLDs was evaluated using the TLD Reader Harshaw, while Gafchromic Film EBT2 was scanned using Epson Perfection V700 scanner with transmission mode, red channel and resolution 72 dpi. The point dose measurements were compared between planned dose TPS Eclipse ver. 11 and measured dose at target volume organ and organ at risk for IMRT and VMAT techniques. The result is the dose difference at target volume for IMRT and VMAT are less than 5%. Similarly, the dose difference at Bladder and Rectum for both techniques are also less than 5% due to the position of OAR is very close to target volume. On the other hand, the dose difference at Femoral head are more than 5% for both techniques because the location of OAR already in low gradient dose. Furthermore, the difference dose of the target volume for IMRT technique is tends to be smaller than VMAT either for TLD and film detectors. The dose difference at point dose of target volume between IMRT and VMAT techniqe are less than 1% but it occur in random number for organ at risk. More over, the surface dose of IMRT tend to be smaller than VMAT dose if we are using TLDs, but the dose of EBT2 films tend to be similar between IMRT and VMAT techniques, Have been done the dose verification of the target and healthy tissues around by placing the TLD Rod LiF100 and EBT2 Gafchromic film at slab hole of pelvic part of the Alderson Rando phantom for prostate cancer simulation. The Exposed TLDs was evaluated using the TLD Reader Harshaw, while Gafchromic Film EBT2 was scanned using Epson Perfection V700 scanner with transmission mode, red channel and resolution 72 dpi. The point dose measurements were compared between planned dose TPS Eclipse ver. 11 and measured dose at target volume organ and organ at risk for IMRT and VMAT techniques. The result is the dose difference at target volume for IMRT and VMAT are less than 5%. Similarly, the dose difference at Bladder and Rectum for both techniques are also less than 5% due to the position of OAR is very close to target volume. On the other hand, the dose difference at Femoral head are more than 5% for both techniques because the location of OAR already in low gradient dose. Furthermore, the difference dose of the target volume for IMRT technique is tends to be smaller than VMAT either for TLD and film detectors. The dose difference at point dose of target volume between IMRT and VMAT techniqe are less than 1% but it occur in random number for organ at risk. More over, the surface dose of IMRT tend to be smaller than VMAT dose if we are using TLDs, but the dose of EBT2 films tend to be similar between IMRT and VMAT techniques]"
2015
T43792
UI - Tesis Membership  Universitas Indonesia Library